75
Config
Turn ON/OFF the Sense Function
When the DUT outputs large currents, the Sense terminal can be used to
accurately test the voltage across the output terminals to compensate for the
voltage drop on the load. Set
Vrmt to ON to enable the remote sense test
function.
Turn ON/OFF SOF Function
SOF stands for Stop On Fail. When enabled, the auto test function will stop
running immediately when any step
fails during an auto test. If SOF is
disabled, the auto test will continue to run until the test is completed, even if a
failure is detected.
Break-Over Voltage
When testing power products with slow voltage rise rates, if the electronic load
input is opened before power, the power may latch protection. In this way, the
user may set VON value. The electronic load only latches when the power
voltage is higher than this value.
Turn ON/OFF Von-Latch Function
The Von Latch function locks the input voltage to the Von-Latch value if the Von-
Latch value is measured. The load will start testing when the the power output
voltage rises and is higher than the break-over voltage. When the power voltage
drops and is lower than the break-over voltage, the load will continue to latch at
the Von-Latch value, as shown in figure 3-11.