23
polarity of the load to the (-) terminal of the channel output. Incorrect
polarity may cause damage to the instrument or the DUT.
2. Press CP to enter the main interface of CP mode, as shown in Figure
2-12.
Figure 2-12 CP Mode Main Interface
3. Set CP mode current range (5 A or 30 A) and voltage range (36 V or 150
V).
4. Set power value.
5. Press On/Off to turn on the channel input. At this time, the actual input
voltage, current, resistance and power will be displayed on the main
interface.
Note:
The load begins to sink power only if the input voltage is greater than
the conduction voltage of the system (default value is 0 V)
Warning
To avoid electric shock, ensure that the DUT is connected to the input
terminals of the load properly before you turn on the channel input.
6. Press the Display key to enter the waveform display interface, as shown
in Figure 2-13. By default, the power waveform is displayed. When the
input voltage changes, the load will sink a constant power. Press the
Display key again to exit the waveform display interface and return to the
main interface of CP mode.