83
Figure 3-16 SLTM tab
3. Input the low voltage and high voltage, then turn on the TMon
4. Connect the load to the DC power under test. Set the power output higher
than the V_High. At this time, the power is OFF.
5. Set a sink current of the load in CC mode, then turn on the load input.
6. Turn on the power under test, the T_Rise is displayed quickly on the LCD
screen which is the voltage rise time.
7. Turn off the power under test, the T_Fall on the LCD screen is the falling
time.
8. Close all instruments and finish the test.
Limits
I_Protect
T3EL15030xP electronic load can have overcurrent protection set by the
electronic load control software. Turn on the load input, if the sink current of the
load is higher than the protection current, the load will turn off it’s input
automatically and pop-up an overcurrent protection message after a delay
period.
Operating Steps
1. Press Shift and CP key to enter system utility function interface. Then enter
“Config” tab and select “I_Protect” tab to enter I_Protect function.
2. Turn on the “TMon” tab to enable the I_Protect. Then set the protection
current and delay time, as shown in Figure 3-17.