18
Constant voltage mode
Figure 2-5 Voltage-Current Relationship Schema under CV Mode
Operating Steps
1. Turn off the instrument, as shown in Figure 2-2, connect the DUT
and the channel input terminals on the front panel of the load.
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. Incorrect polarity may cause damage to the instrument or
the DUT.
2. Press CV to enter the main interface of CV mode, as shown in
Figure 2-6.
Figure 2-6 CV Mode Main Interface
3. Set CV mode current range (5 A or 30 A) and voltage range (36 V