28
Figure 2-18 Waveform display interface of Cont mode
Pulse mode
Dynamic test operations using pulse mode configures the load to source the
low value (Level A) until a valid trigger is received. At this time, the load settings
will change to the B values. The settings will switch back to the A values after
maintaining B for the set pulse width time, as shown in Figure 2-19.
Figure 2-19 Transient CC Pulse Mode(Cont)
Take CC mode as an example (other modes are similar)
Operating Steps
1. Connect device
Power on the instrument, connect the DUT and the channel input
terminals of the electronic load, as shown in Figure 2-2.
2. Set running parameters
Press the Shift + CC key on the front panel to enter transient test
operation. CC continuous mode is not only the default mode but also