stops discharging automatically. The default unit for C_Stop is
milliamp-hour (mAh).
Set T_Stop
The discharge time in the Battery test function. When the accumulated
time is greater than the cut-off time, the load stops discharging
automatically. The default unit for T_Stop is seconds (s).
3.
Turn on the input channel
Press On/Off to turn on the channel input. At this time, the actual
discharge voltage, current, time and capacity will be displayed on the
main interface. If any of the termination conditions are met, the load will
pop-up a message box stating “Battery test compelete”, which will
disappear by pressing any key.
Note:
Once the channel input is turned on, the load will not start to sink
current until the input voltage is greater than the break-over voltage.
4. View waveform
Press the Display key to enter the waveform display interface, as
shown in figure 2-40. By observing the CC waveform with the Data
parameter set to current (I), you can see
the load discharge with a
constant current. Press the Display key again to exit the waveform
display interface and return to the main interface of OPPT test mode.
Figure 2-40 Waveform display interface of BAT