33
The parameters for the pulse operation mode include Function, Mode,
range, A_ Level, B_ Level, rising slew rate, falling slew rate and trigger
selection. The interface menu can be divided into three pages.
Set range
Current range: 5 A or 30 A
Voltage range: 36 V or 150 V
Set A_level
The sink current toggles between a high value and a low value in Pul
mode. A_Level indicates the low value. The default unit for A_Level is
Ampere (A).
Set B_level
The sink current toggles between a high value and a low value in Pul
mode. B_Level indicates the high value. The default unit for B_Level is
Ampere (A).
Slew_RIS and Slew_FALL
Slew_RIS and Slew_FALL: The rate of change to a new level from the
current sink value of the load. Its unit is A/us and the setting range is
0.001 - 2.5 A/us.
3. Enable trigger
Press Shift + CP key to enter Utility interface .After pressing “Config”
key the trigger source can be set in page 2 of the menu. There are
three trigger types:Manual/Ext/Bus. Manual displays a trigger softkey
on page 3 of the Dynamic test. Pressing this key will trigger the
step. Ext sets the trigger type to external, which looks for a valid
trigger on the rear panel Ext Trig input. Bus is a trigger sent over
the remote control bus.
See “Set trigger” in Section 3 for more details.
Press On/Off to turn on the channel input. At this time, the actual
input voltage, current, resistance and power will be displayed on the
main interface.