41
will pop-up a message box stating “Test complete”, the load will then
disable the output. If the OPPT test fails, the interface of the load will
pop-up a message box stating “Below limit, Test fail”, “Over limit,
Test fail”, etc. then the input channel of the load will be switched off
automatically.
4. View waveform
Press Display key to enter waveform display interface, as shown in
figure 2-36. The sink power will have a stepped increase in the
waveform after setting the Data selection to be power (“P”). Press the
Display key again to exit the waveform display interface and return to
the main interface of OPPT test mode.
Figure 2-36 Waveform display interface of OPPT
Battery Test Function
The Battery test function can be used in CC,CP, or CR mode. The primary
termination conditions for the Battery test function include: cut-off voltage, cut-off
capacity, and/or discharge time. When any one of the three conditions is met, the
load immediately stops discharging. When only one or two condition(s)
is/are selected as the termination condition(s) for the battery test, please set the
unused termination conditions to the "OFF" state. The C-Add function can
record accumulated capacity if required.