User Manual
77
Figure 35 State DropOut Trigger
Runt Trigger
The Runt trigger looks for pulses that cross one threshold but not another as shown in the
picture below.
Positive runt pulse
High level
Low level
Negative runt pulse
●
A positive runt pulse crosses through a lower threshold but not an upper threshold.
●
A negative runt pulse crosses through an upper threshold but not a lower threshold.
To trigger on runt pulse:
1. Press the
Setup
button on the front panel to enter the
TRIGGER
system function
menu.
2. Press the
Type
softkey; then turn the
Universal Knob
to select
DropOut
and push
down the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as the
trigger source.
4. Press the
Polarity
softkey to select
Positive
or
Negative
pulse to trigger.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...