User Manual
165
To Set and Perform Pass/Fail Test
Do the following steps to set and perform pass/fail test:
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
2. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
3. Press the
Pass/Fail
softkey to enter the PASS/FAIL function menu.
4. Press the
Enable Test
softkey to select
On
to enable the pass/ fail test.
5. Press the
Source
softkey to select the desired channel.
6. Press the
Mask Setting
softkey to enter the MASK function menu.
7. Press the
X Mask
or
Y Mask
softkey; and then turn the
Universal Knob to
select the
desired value. The range is from 0.02div to 4div.
8. Press the
Create Mask
button to create the mask. Whenever the
Create Mask
softkey is pressed the old mask is erased and a new mask is created.
9. Press the
UP
softkey to return to the PASS/FAIL function menu.
10. Press the
Msg Display
to select
On
or
Off
to turn on or off the message display.
When
On
is selected, the test result will be displayed in the upper-right message box
of the screen.
●
Fail:
the number of waveforms that failed to match the mask (mask violation).
●
Pass:
the number of waveforms within the mask.
●
Total:
the total number of waveforms that have been acquired. The total of
Pass
and
Fai
l.
11. Press the Next Page softkey to go to the second page of the PASS/FAIL function
menu.
12. Press the Fail softkey to select On or Off to turn on or off the function.
●
On:
when failed waveforms are detected, the oscilloscope will stop the test and
enter the STOP state. At this point, the results of the test remain on the screen
(if the screen is turned on) and only one pulse is output from the
[TRIG OUT]
connector (if enabled) on the rear panel.
●
Off:
the oscilloscope will continue with the test even though failed waveforms are
detected. The test results on the screen will update continuously and a pulse will
be output from the
[Trigger Out]
connector at the rear panel each time a failed
waveform is detected.
13. Press the Output softkey to turn on or off the sound.
●
: When the failed waveforms are detected, the beeper sounds.
●
: The beeper will not sound even if the failed waveforms are detected.
14. Press the Next Page softkey to return to the first page of the PASS/FAIL function
menu.
15. Press the Operate softkey to perform the test.
●
: Current state is stop; press the softkey to start the pass/fail test.
●
: Current state is running; press the softkey to stop the pass/fail test.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...