User Manual
197
The table below shows the maximal number of acquisition frames according to
the sampling rate and current number of sampling points.
Sample Rate
Curr (pts)
Max. Frame
Sample Rate
Curr (pts)
Max. Frame
1GSa/s
≦
280
80000
500MSa/s
35K
783
700
57227
70K
391
1.4K
33528
140K
195
2.8K
18338
350K
77
7K
7773
700K
38
14K
3982
1.4M
18
28K
1993
3.5M
6
70K
798
7M
3
140K
398
14M
1
280K
198
≤
250MSa/s
3.5K
3779
700K
78
7K
1891
1.4M
38
14K
945
2.8M
18
17.5K
757
7M
7
35K
378
14M
3
70K
188
500MSa/s
≦
35
80000
140K
93
70
77026
175K
74
140
65667
700K
17
350
45526
1.4M
8
700
29140
1.8M
6
1.4K
16945
3.5M
3
3.5K
7510
7M
1
7K
3912
14M
1
14K
1958
4. Press the
softkey to replay the waveform from the current frame to 1.
5. Press the
softkey to stop replay.
6. Press the
softkey to replay the waveform from the current frame to the last
frame.
●
The maximal acquisition number is determined by the current sampling point
(
Curr
value) and sampling rate.
●
The History Mode maximum acquisition count can only be reached if enough
acquisitions have been acquired, otherwise a smaller number of acquisitions
will be shown.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...