180 User Manual
To Update Firmware and Configuration
The firmware and configuration can be updated directly via USB flash drive.
Do the following steps to update the firmware:
1. Insert the USB flash drive which contains the firmware and the configure files into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Firmware
softkey and the message “
Press “Single” to continue and
Press “Run/Stop” to exit!
7. Press the
Single
button to enter the SAVE/RECALL file system.
8. Turn the
Universal Knob
to select the update file which should be with an ADS postfix,
and then press the
Load
softkey to start the firmware update. The process needs about
7 minutes. And during the update, do not turn off the oscilloscope or disconnect the
oscilloscope power, otherwise the oscilloscope will not start again.
9. After the update has finished, the screen will display the message “
Update success,
please restart the DSO
”, and the oscilloscope buzzer will sound.
10. Restart the oscilloscope to finish the firmware update.
Do the following steps to update the configuration:
1. Insert the USB flash drive which contains the firmware and the configure files into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Configure
softkey to enter the SAVE/RECALL file system.
7. Turn the
Universal Knob
to select the update file which will have a CFG postfix,
and then press the
Load
softkey to start updating the firmware. The process
needs about 30 seconds.
8. After the update has finished, the screen will display the message “
Update success,
Please restart the DSO
”, and the oscilloscope buzzer will sound.
9. Restart the oscilloscope to finish the configuration update.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...