Adjust Trigger Delay
Turn the Position Knob on the front panel to adjust the trigger delay of the waveform.
During the modification, waveforms of all the channels would move left or right and the
trigger delay message at the upper-right corner of the screen would change accordingly.
Press down this knob to quickly reset the trigger delay tothe center of the screen.
Changing the delay time moves the trigger point (solid inverted triangle) horizontally and
indicates how far it is from the time reference point. These reference points are indicated
along the top of the display grid.
All events displayed left of the trigger point happened before the trigger occurred. These
events are called pre- trigger information, and they show events that led up to the trigger
point.
Everything to the right of the trigger point is called post- trigger information. The amount of
delay range (pre- trigger and post- trigger information) available depends on the time/div
selected and memory depth.
The position knob works (in the Normal time mode) while acquisitions are running or when
they are stopped.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...