Note: Because differentiation is very sensitive to noise, it is helpful to set acquisition
mode to
Average.
Figure 60 Differential Function Operation
You can use differentiate to measure the instantaneous slope of a waveform. For example,
the slew rate of an operational amplifier may be measured using the differentiate function.
Integrate
dt (integrate) calculates the integral of the selected source. You can use integrate to
calculate the energy of a pulse in volt- seconds or measure the area under a waveform.
dt plots the integral of the source using the "Trapezoidal Rule". The equation is:
I
n
�
c
0
�
∆
t
∑
i 0
n
y
i
Where:
●
I = integrated waveform;
●
Δt = point- to- point time difference;
●
y = channel 1, 2, or REFA,REFB
●
c
o
= arbitrary constant;
●
i = data point index;
The integrate operator provides an
Offset
softkey that lets you enter a DC offset
correction factor for the input signal. Small DC offset in the integrate function input (or
even small oscilloscope calibration errors) can cause the integrate function output to
"ramp" up or down. This DC offset correction lets you level the integrate waveform.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...