86 User Manual
●
10 Bit Address & Data
— the oscilloscope will be triggered when the
following conditions are satisfied.
➢
The address’s length must be 10 bits and the address’s value is the same as
the address set value.
➢
If you have set either Data1’s or Data2’s value, and the signal has data that
is the same as those values in that order.
(
Note
: If the set value is 0xXX, any data value will be matched)
Write
●
Data Length
— When SDA data’s length is equal to the value of Byte Length and
address’s length is the same as the set value, the oscilloscope will be triggered.
Byte length is in the range of 1 to 12 bits.
Operation steps:
1.
Press
Setup
to enter the
TRIGGER
function menu.
2.
Press
Type
and select
Serial
.
3.
Press
Protocol
and select
I2C
.
4.
Press
Trigger Setting
softkey.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...