122 User Manual
To display an FFT waveform:
1. Press the
Math
button on the front panel to open the MATH function menu.
2. Press the
Operation
softkey and then turn the
Universal Knob
to select
FFT
. The
resulting math waveform is displayed in white and labeled with “M”.
3. Press the
Source
softkey, and then turn the
Universal Knob
to select the source to
do an FFT operation. Analog channels can be used as the source.
4. Press the
Window
softkey, and then turn the
Universal Knob
to select an appropriate
window.
Spectral leakage can be considerably decreased when a window function is used. The
oscilloscope provides five windows (Rectangle, Blackman, Hanning, Hamming and
Flattop) which have different characteristics and are applicable to measure different
waveforms. You need to select the window function according to different waveforms
and their characteristics. Please read the table below carefully to make an appropriate
option choice according to the input signal.
Window
Applications and Characteristics
Rectangle
This is normally used when the signal is transient (completely
contained in the time-domain window) or known to have a fundamental
frequency component that is an integer multiple of the fundamental
frequency of the window. Signals other than these types will show varying
amounts of spectral leakage and scallop loss, which can be corrected by
selecting another type of window
Hanning
Reduced leakage and improve amplitude accuracy.
Frequency resolution is also reduced.
Hamming
Reduce leakage and improve amplitude accuracy.
Frequency resolution is also reduced.
Flat Top
This window provides excellent amplitude accuracy with moderate
reduction of leakage, but with reduced frequency resolution.
Blackman
Leakage is reduced to a minimum, but with reduced frequency
resolution.
5. Press the
Center
softkey, and then turn the
Universal Knob
to select the desired
center frequency.
6. Press the
Hz/div
softkey, and then turn the
Universal Knob
to select the desired
resolution frequency.
7. Press the
Scale
softkey, and then turn the
Universal Knob
to select the desired
vertical FFT scale
8. Press the
Ref Level
softkey, and then turn the
Universal Knob
to select the desired
vertical FFT offset.
9. Press the
Unit/Load
softkey to enter Unit/Load menu. Press
Unit
softkey to select the
unit of vertical axis. The units of the vertical axis can be dBVrms, dBm or Vrms which
use a logarithmic scale or a linear scale to display vertical amplitude respectively. If
you need to display the FFT frequency spectrum in a relatively larger dynamic range,
dBVrms is recommended. Press
Ext Load
softkey, and then turn the
Universal Knob
to select the external load value.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...