User Manual
69
Figure 30 Video Trigger
To Use Custom Video Trigger
Custom video trigger supports frame rate of 25Hz, 30Hz, 50Hz and 60Hz, and the line
range is available from 300 to 2000. The steps below show how to set custom trigger.
1. Press the
Setup
button on the front panel to enter the TRIGGER function menu.
2. Press the
Type
softkey; then use the
Universal Knob
to select
Video
and push down
the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select
CH1
as the trigger
source, and push down the knob to confirm.
4. Press the
Standard
softkey; turn the
Universal Knob
to select
Custom
, and push
down the knob to confirm.
5. Press the
Setting
softkey to enter the custom setting function menu. Press the
Interlace
softkey; turn the
Universal Knob
to select the desired value.
6. Press the
Of Field
softkey; turn the
Universal Knob
to select the desired value.
7. Press the
Sync
softkey to enter the TRIG ON menu to set the line and field.
●
Press the Type softkey to select
Select
or
Any
.
●
If the
Type
option set to
Select
, press the Line softkey; turn the
Universal Knob
to select the desired value. Press the
Field
softkey; turn the
Universal Knob
to
select the desired value.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...