202 User Manual
Troubleshoo�ng
The commonly encountered failures and their solutions are listed below. Please contact
Teledyne Test Tools
if the problem is not listed below and you need further assistance.
1.
The screen has no display after power on:
1)
Check whether the power is correctly connected.
2)
Check whether the power switch is on.
3)
Check whether the fuse has blown. If the fuse needs to be changed, please
use the specified fuse.
4)
Restart the instrument after finishing the above checks.
5)
If it still does not work correctly, please contact
Teledyne Test Tools.
2.
The signal is sampled but no waveform is displayed:
1)
Check whether the probe is correctly connected to the DUT waveform.
2)
Check whether the probe or BNC cable is correctly connected to the oscilloscope
BNC.
3)
Check whether the oscilloscope channel is turned on.
4)
Check whether there are signals generated from the item to be tested (you can
connect the probe compensation signal to the oscilloscope channel to ensure
that the scope channel and probe are working correctly).
5)
Resample the signal, press Run.
3.
The tested voltage amplitude is greater or lower than the actual value (note that
this problem usually occurs when a probe is used):
Check whether the attenuation coefficient of the channel complies with the
attenuation ratio of the probe.
4.
There is an unstable waveform display:
1)
Check that the trigger signal source has a suitable signal to trigger on.
2)
Check the trigger type: general signals should use the “Edge” trigger.
3)
Change the trigger holdoff setting if the signal is a burst type waveform.
5.
No display after pressing Run/Stop:
Check whether the trigger mode (TRIGGER) is on “Normal” or “Single” and whether
the trigger level exceeds the waveform voltage range. Try setting the trigger level to
the middle of the expected waveform voltage range, or set the mode to “Auto”.
Note: using
AUTO
will automatically trigger the scope whether there is a waveform
present or not.
Содержание T3DSO1000
Страница 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Страница 21: ...User Manual 3 Appearance and Dimensions Figure 1 Front View T3DSO1102 scope Figure 2 Top View T3DSO1102 scope...
Страница 22: ...4 User Manual Figure 3 Front View All scopes except T3DSO1102 Figure 4 Top View All scopes except T3DSO1102...
Страница 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 90: ...72 User Manual Figure 32 Relative Window Trigger...
Страница 92: ...74 User Manual Figure 33 Interval Trigger...
Страница 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Страница 118: ...100 User Manual Figure 50 UART Trigger...
Страница 145: ...User Manual 127 function output Figure 63 Square Root...
Страница 205: ...User Manual 187 Figure 100 Option Information...