P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 53
13
ST_FLOW_ALARM
YES
Status Bit - Flow alarm
Logic High = Sample flow out of spec
Logic Low = Flows within spec
14
ST_TEMP_ALARM
YES
Status Bit - Temperature alarm
Logic High = One or more temps out
of spec
Logic Low = Temps within spec
15
ST_DIAG_MODE
YES
Status Bit - In Diagnostic mode
Logic High = M100A in Diagnostic
mode
Logic Low = Not in Diag mode
16
ST_POWER_OK
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
17
ST_PRESS_ALARM
YES
Status Bit - Flow alarm
Logic High = Sample pressure out of
spec
Logic Low = pressure within spec
18
ST_HIGH_RANGE
YES
Status Bit - Autorange High Range
Logic High = M300 in high range
Logic Low = M300 in low range
19
ST_SYSTEM_OK
YES
Status Bit - System OK
Logic High = No instrument warning
present
Logic Low = 1 or more alarm present
20
ST_BENCH_ALARM
YES
Status Bit - Bench Temperature Alarm
Logic High = Bench Temp out of spec
Logic Low = Bench Temp in spec
21
ST_SOURCE_ALARM
YES
Status Bit - IR Source Alarm
Logic High = IR Source output too
low
Logic Low = IR Source output
normal
22 ST_WHEEL_ALARM YES Status
Bit - Wheel Temperature Alarm
Logic High = Wheel Temp out of spec
Logic Low = Wheel Temp in spec
23 CO_MEASURE NO
IR
detector reading during measure
phase. Typically 2500-4500 mV
24 CO_REFERENCE NO IR
detector reading during reference
phase. Typically 2500-4500 mV
25
SAMPLE_PRES
NO
Sample pressure in mV. Typical sea
level value = 4300mV for 29.9" HG-
A.
26
SAMPLE_FLOW
NO
Sample flow in mV.
27
SAMPLE_TEMP
NO
Sample temp in mV
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...