P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 51
The
TEST
button is used to scroll through the test measurements until the one of interest
is displayed. To turn the test on press the
OFF/ON
button. Viewing test measurements
in the diagnostic mode is especially useful for viewing the results of a diagnostic test.
The
PREV
button goes to the previous diagnostic test. When pressed, the CPU turns the
current diagnostic test OFF if it is ON. The
NEXT
button goes to the next diagnostic test.
When pressed, the CPU turns the current diagnostic test OFF if it is ON. The
EXIT
button exits the diagnostic mode and turns all the diagnostic tests OFF. This ensures that
a diagnostic test is not accidentally left ON. A message is also sent to the RS-232
channel to indicate that the diagnostic mode has been exited.
DIAGNOSTIC TESTS
Test #
Name
Signal
Nominal Value
1 NONE
No
output
0
mV
2
CO MEASURE
CO detector measure value
4500 mV
3
CO REFERENCE
CO detector reference
value
4200 mV
4
SAMP PRESS
Sample pressure
125 mV
5
SAMP FLOW
Sample flow
3000 mV
6
SAMP TEMP
Sample temperature
3000 mV
7
BENCH TEMP
Optical bench temperature
3900 mV
8
WHEEL TEMP
Outputs filter wheel temp.
3900 mV
9
CHASSIS TEMP
Outputs Chassis temp
2740 mV
10
DCPS VOLT
Outputs DC power
2500 mV
11
DAS AVERAGE
Current DAS average
Any
TABLE 5.1
5.2.1 Signal I/O
The signal I/O diagnostic mode gives the user access to the digital and analog inputs and
outputs on the V/F board. The digital outputs can be controlled through the keyboard.
Any signals manually changed through the signal I/O menu will remain in effect until
you leave the signal I/O menu. At that time the analyzer will regain control of these
signals. To enter the signal I/O test mode, press
SETUP-MORE-DIAG-ENTR
. When
the diagnostic mode is entered, a message is sent to the RS-232 channel indicating entry
into the diagnostic mode. Use the
PREV
and
NEXT
buttons to scroll through the signals.
Edit buttons will appear for the signals that can be controlled by the user. Press
JUMP
to
skip to a specific I/O Signal.
Table 5.2 below lists the I/O signals available for the M300:
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...