P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 36
3.7 Remote zero/span check or adjustment
(RS-232)
Besides Z/S checking from the front panel, automatic Z/S checking, and remote Z/S
checking via the contact closure inputs, the Analyzer can also be checked via the RS-232
interface. Remote checking via the RS-232 interface supports zero, span, and zero
followed by span check, and is identical to remote check via the contact closure inputs.
This RS-232 control feature is provided mainly so that a host computer at another
location can control the Analyzer. See Section 7.0 for detailed information on using the
RS-232 interface to do a remote Z/S check.
3.8 Power-on hold off
Whenever the Model 300 is powered on it will go through a HOLD-OFF sequence (see
Section 3.10 below) like it does after a zero/span check.
3.9 Hold off
Every type of check or adjustment (zero, span, manual, remote, etc.) is followed by a
hold off period of from 1 to 20 minutes, during which time the internal data acquisition
system (DAS) does not accumulate CO readings into the DAS average. To set the hold
off time, press
SETUP-MORE-VARS
and press
NEXT
until the variable
HOLDOFF_TIME is displayed. To change the setting, enter a number from 1 to 20 and
then press
ENTR
to store the new value or
EXIT
to leave the hold off time unchanged.
NOTE
REMOTE ZERO AND SPAN ADJUSTMENT ALTER THE
FORMULAS USED TO COMPUTE THE CARBON
MONOXIDE READINGS IF DYN_ZERO OR DYN_SPAN ARE
ENABLED. THIS METHOD OF CALIBRATION IS NOT
APPROVED BY USEPA AND IS NOT INTENDED TO
REPLACE THE USEPA APPROVED CALIBRATION.
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...