P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 96
the front panel. It should read the same as the version number that was located on
the top right corner of the label on the PROM.
8.
Re-enter any non-default settings such as RANGE or ASEQ. Re-enter the CONC
value in the CAL menu. Check all settings to make sure that expected setup
parameters are present.
9.
Re-calibrate the analyzer so that the default slope and intercept are overwritten
with the correct values.