P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 6
7.
Turn on the M300 by switching the switch on the lower right corner of the
front panel (See Figure 2.1). The front panel display should light with a
sequence of messages, -API - M300 - software version number, then a
normal display as shown in Figure 2.2.
8.
Allow about 60 minutes for the temperatures to come up to their
respective setpoints then press the left most button on the front keyboard
to scroll through the TEST values. Compare these values to those noted
during the final factory checkout listed in Table 1.1. The values observed
should closely match the Table 1.1 values.
9.
Select the range on which the analyzer will be calibrated.
a. From the SAMPLE menu press
SETUP
to enter the SETUP menu.
(See Figure 2.2 for appearance of front panel.)
b. Enter the PASSWORD (818).
c. Press
RNGE
(RANGE).
d. Press
SET
.
e.
Enter the derived full scale range for analog outputs and press
ENTR.
f. Press EXIT 2 times to return to the SAMPLE menu.
10.
Adjust the analyzer zero point.
a.
Input zero air into the sample port.
b. Press
CAL
from the SAMPLE menu and enter the password
(818)
.
c. Press
ZERO.
d.
It usually takes about 5 to 10 minutes for the reading to stabilize
near zero. After a stable reading has been obtained press
ENTR
.
The display should now read 0.0 ppm carbon monoxide.
CAUTION
CHECK THAT ANALYZER IS SET UP FOR PROPER
VOLTAGE AND FREQUENCY.
CAUTION
POWER PLUG MUST HAVE GROUND LUG.
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...