P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 33
Range To Cal:
This setup parameter is enabled only if the range mode is set to Dual or
Auto. This parameter determines which range the sequence will check.
Examples of possible sequences are as following under any one of three available SEQx.
Example 1
: to perform a 15 minute zero check once per day at 10:30 pm, 12/20/93.
1. MODE: ZERO
2. TIMER ENABLED: ON
3. STARTING DATE: 12/20/93
4. STARTING TIME: 22:30
5. DELTA DAYS: 1
6. DELTA TIME: 00:00
7. DURATION: 15
8. CALIBRATE: OFF
Example 2
: to perform a 15 min zero span check once per day retarding 15 minutes
everyday starting at 11:30 pm, 12/20/93.
1. MODE: ZERO-SPAN
2. TIMER ENABLED: ON
3. STARTING DATE: 12/20/93
4. STARTING TIME: 23:30
5. DELTA DAYS: 0
6. DELTA TIME: 23:45
7. DURATION: 15
8. CALIBRATE: OFF
Example 3
: to perform span check once per week starting at 11:30 pm, 12/20/93
1. MODE: SPAN
2. TIMER ENABLED: ON
3. STARTING DATE: 12/20/93
NOTE
THE CALIBRATE FEATURE OF AUTO SEQUENCES
ALTERS THE FORMULAS USED TO COMPUTE THE
CARBON MONOXIDE READING. THIS METHOD OF
CALIBRATION IS NOT APPROVED BY USEPA AND IS
NOT INTENDED TO REPLACE THE USEPA APPROVED
CALIBRATION METHOD.
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...