P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 48
Step
Action
Comment
1.
Press
SETUP-DAS-
EDIT
Enter DAS menu to edit Data
Channels
2.
Press
PREV/NEXT
Select Data Channel to edit
3.
Press
EDIT
Edit selected Data Channel
4.
Press
SET>
(5 times)
Scroll through setup properties until
RS-232 REPORT: OFF is displayed
5.
Press
EDIT
Edit selected setup property
6.
Toggle
OFF
to
ON
Change RS-232 REPORT property
7.
Press
ENTR
Accepts change
8.
Press
EXIT
(4 times)
Exits back to sample menu
See Section 7.3.1 for more information on DAS reporting through the RS-232 interface.
4.12 Software configuration
The software configuration can be displayed by entering the button sequence
SETUP-CFG-LIST
. For example the M300 could display:
"CO MACHINE"
"SBC40 CPU"
Stating that the instrument was an Carbon Monoxide Analyzer using the SBC40
computer. This feature is useful for showing any special features that are present in the
currently installed PROM.
4.13 Summary of setup functions
The setup functions are summarized in Table 4.1 in terms of the button sequences used to
access them.
SETUP FUNCTIONS
Button Sequence
Function
Default
Limits
SETUP-CFG-LIST List
Software
Configuration
CO Mach
N/A
SETUP-ACAL-MODE Define/Change
AutoCal Sequences
Disabled Zero,
Span,
Zero-Span
SETUP-DAS-EDIT Define/Change
DAS
Data Channels
SETUP-DAS-VIEW View
DAS
Data
N/A
N/A
SETUP-RNGE-MODE Set
Range
Mode
Single
Single,
Dual, Auto
SETUP-RNGE-SET
Set D/A output range
50 ppm
1-1000ppm
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...