P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 30
3.1.3 Dual Range Calibration
If the analyzer is being operated in Dual Range mode or Auto-Ranging mode, then the
High and Low ranges must be independently calibrated. When the analyzer is in Dual or
Auto Range mode you will be prompted to enter the range to calibrate whenever you
enter a calibration command from the front panel. Press
HIGH
or
LOW
followed by the
ENTR
button to proceed with the calibration. To calibrate the other range you must exit
to the sample menu and restart the calibration. See Section 4.6 for more information on
the Range Modes. The following procedure shows an example of how to calibrate the
two ranges with calibration gas coming in through the sample port:
Step
Action
Comment
1. Press
CAL
Analyzer enters M-P calibration mode. Calibration gas
source should be set to deliver zero gas to the sample port.
2. Press
LOW-ENTR
Select range to calibrate
3.
Wait 15 min.
Wait for CO reading to stabilize at zero value.
4. Press
ZERO-ENTR
Changes
calibration
equations for Low range so analyzer
will read zero.
5. Press
CONC
Enter span gas concentration for Low range.
6.
Key in span
concentration
Enter span gas concentration for Low Range. Set
calibration gas source to deliver span concentration
7. Press
ENTR
8.
Wait 15 min.
Wait for CO reading to stabilize at span value
9. Press
SPAN-ENTR
Changes
calibration
equations for Low range so analyzer
will read span value.
10. Press
EXIT
Exits back to sample menu
Repeat steps 1-10 for High range.
3.2 IZS zero/span check (Option)
If the IZS (Internal Zero Scrubber) option has been installed and a source of CO span gas
has been connected, the operator can check the zero and span setpoints of the analyzer at
any time by pressing the
CALZ
or
CALS
button.
Pressing
CALZ
switches the Sample/Cal valve and allows the analyzer to draw air
through the zero air scrubber. After a few minutes the CO reading should go to zero. If
it doesn't, the operator may press the
ZERO
button followed by
ENTR
.
CALS
works like
CALZ
except that externally supplied span gas is drawn through the
analyzer. After a few minutes the CO reading should approach the span level (typically
40 PPM). If it doesn't, the operator may press the
SPAN
button, followed by
ENTR
.
The expected span gas concentration may be changed by pressing the
CONC
button.
The operator can exit the IZS calibration only by pressing the
EXIT
button. If either the
ZERO
or
SPAN
buttons were pressed and
ENTR
was not pressed, the Analyzer will
beep once to indicate that no changes have been made.
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...