P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 94
b.
Open the front panel and remove the transparent filter cover and knurled
retaining ring.
c.
Remove the teflon hold-down O-ring.
d.
Remove the old filter element and discard.
e.
Install a new filter element in the filter cavity. Be careful with the
element, it is fragile.
f.
Replace the hold-down O-ring on top of the filter element.
g.
Replace the filter top and re-tighten.
h. Leak
check.
i.
Turn on the Analyzer.
11.3 Leak checking
There are two methods of leak checking:
11.3.1 Using a leak checker
Turn the power off. Disconnect the fittings from the pump and bypass the pump.
Connect the leak checker to the sample inlet of the Model 300. Cap the exhaust of the
analyzer. Set the leak checker to pressure mode.
Leave the checker on until 15 psi is achieved. Close the valve and ensure the pressure
remains at 15 psi for at least 5 minutes. If pressure drops more than 1 psi, there is a leak
and it must be repaired.
If there is a leak present, pressurize the Model 300 to 15 psi and put soap bubble
solution on pneumatic assemblies until the leak is found.
CAUTION
DO NOT EXCEED 15 PSI OF PRESSURE.
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...