P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 52
#
Signal
Contro
l
Description
0
DISP_BROWNOUT
NO
Display brownout is used to keep the
display from getting corrupted during
low line voltage conditions. Circuitry
on the Power Supply board (00015)
senses low line voltage and sets this
bit. The CPU reads this and generates
the BROWNOUT_RST signal
described below.
1
EXT_ZERO_CAL
NO
Shows state of status input bit to cause
the M300 to enter Zero Calibration
mode. Use to check external contact
closure circuitry.
2
EXT_SPAN_CAL
NO
Shows state of status input bit to cause
the M300 to enter the Span Calibration
mode. Use to check external contact
closure circuitry.
3
SYNC_OK
NO
Indicates that demodulation circuitry
on the Sync/Demod Board (00798) is
able to lock-in on the detector signal.
4
SPAN_VALVE
YES
Switches the Zero/Span valve. Use
this bit to test the valve function.
5 CAL_VALVE YES
Switches
the
Sample/Cal valve. Use
this bit to test the valve function.
6
BENCH_HTR
YES
Shows the status of the optical bench
heater. This has the same function as
the LED in the power supply module.
7
WHEEL_HTR
YES
Shows the status of the filter wheel
heater. This has the same function as
the LED in the power supply module.
8
DARK_CAL
YES
Turns off the detector input to the
Sync/Demod board for electronics
calibration.
9
ELEC_TEST
YES
Activates the Electric Test diagnostic
circuitry.
10
BROWNOUT_RST
YES Resets the DISP_BROWNOUT
circuitry described above.
11
ST_ZERO_CAL
YES
Status Bit - Zero Calibration mode
Logic high = M300 in Zero cal mode
Logic low = Not in Zero cal mode
12
ST_SPAN_CAL
YES
Status Bit - Span Calibration mode
Logic high = M300 in Span cal mode
Logic low = Not in Span cal mode
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...