P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 46
Data Parameters
Name
Description
Units
NONE
COMEAS
CO Measure signal
mV
COREF
CO Reference signal
mV
SLOPE1
Slope of Range #1
SLOPE2
Slope of Range #2
OFSET1
Offset of Range #1
mV
OFSET2
Offset of Range #2
mV
ZSCNC1 Range1
zero/span
measured
concentration
PPM
ZSCNC2 Range2
zero/span
measured
concentration
PPM
CONC1
Concentration of Range #1
PPM
CONC2
Concentration of Range #2
PPM
STABIL Stability
PPM
BENCHTMP
Optical bench temperature
o
C
WHEELTMP
Filter wheel temperature
o
C
SMPFLW Sample
flowrate cc/min
SMPPRS Sample
pressure in-Hg
BOXTMP Box
temperature
o
C
DCPS
DC power supply composite voltage
mV
Example:
Sample the CO concentration (Range #1) once per minute (default) and issue a report
once per 5 minutes, and print the report on the RS-232 channel.
1. Press
SETUP-DAS-EDIT-ENTR-INS
to create a new data channel and begin
editing its properties.
2. Edit the NAME property and give it any name you want, but do not set the blank
name (all dashes) or duplicate names because they won’t be accessible from the RS-
232 interface.
3. For each of the remaining properties, press the SET
>
to view the current setting, and
then press
EDIT
to change the property if necessary. The table below lists the proper
settings for all of the properties for this example:
Содержание 300
Страница 14: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 8 CARBON MONOXIDE ANALYZER FIGURE 1 1 ...
Страница 20: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 14 FLOW DIAGRAM FIGURE 1 3 ...
Страница 22: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 16 REAR PANEL FIGURE 1 5 ...
Страница 23: ...P N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual Page 17 MODEL 300 ASSEMBLY LAYOUT FIGURE 1 6 ...