Teledyne API – Model T300/T300M CO Analyzer
Troubleshooting and Service
249
11.2.
GAS FLOW PROBLEMS
When troubleshooting flow problems, it is a good idea to first confirm that the actual
flow and not the analyzer’s flow sensor and software are in error, or the flow meter is in
error. Use an independent flow meter to perform a flow check as described in Section
10.3.4. If this test shows the flow to be correct, check the pressure sensors as described
in Section 11.5.7.6.
The T300/T300M has one main gas flow path. With the IZS or zero/span valve option
installed, there are several subsidiary paths but none of those are displayed on the front
panel or stored by the DAS.
With the O
2
sensor option installed, third gas flow controlled with a critical flow orifice
is added, but this flow is not measured or reported.
In general, flow problems can be divided into three categories:
1. Flow is too high
2. Flow is greater than zero, but is too low, and/or unstable
3. Flow is zero (no flow)
When troubleshooting flow problems, it is crucial to confirm the actual flow rate without
relying on the analyzer’s flow display. The use of an independent, external flow meter
to perform a flow check as described in Section 10.3.4 is essential.
The flow diagrams found in a variety of locations within this manual depicting the
T300/T300M in its standard configuration and with options installed can help in trouble-
shooting flow problems. For your convenience they are collected here.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...