Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
194
9.2.3.
PERFORMING A BASIC MANUAL CALIBRATION
The following section describes the basic method for manually calibrating the
T300/T300M.
If the analyzer’s reporting range is set for the
AUTO
range mode, a step will appear for
selecting which range is to be calibrated (
LOW
or
HIGH
). Each of these two ranges
MUST
be calibrated separately.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
If the ZERO or SPAN buttons are not displayed during zero or span
calibration, the measured concentration value during this time is out of
the range allowed for a reliable calibration. Refer to Section 10 for
troubleshooting tips.
9.2.3.1.
SETTING THE EXPECTED SPAN GAS CONCENTRATION
Note
When setting expected concentration values, consider impurities in your
span gas.
The expected CO span gas concentration should be 80% of the reporting range of the
instrument (see Section 5.4.1).
The default factory setting is 40 ppm. To set the span gas concentration, press:
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...