Teledyne API - T300/T300M and M300E/EM 04906J DCN7122
APPENDIX A-3: Warnings and Test Functions
A-11
Table A-3: T300/T300M and M300E/EM Test Functions, Revision L.8
TEST FUNCTION NAME
MESSAGE TEXT
DESCRIPTION
RANGE
RANGE=50.0 PPM
3
CO RANGE=50.0 PPM
3, 7
D/A range in single or auto-range modes.
RANGE1
RANGE1=50.0 PPM
3
CO RANGE1=50.0 PPM
3, 7
D/A #1 range in dual range mode.
RANGE2
RANGE2=50.0 PPM
3
CO RANGE2=50.0 PPM
3, 7
D/A #2 range in dual range mode.
CO2RANGE
CO2 RANGE=20 %
7
CO
2
range.
O2RANGE
O2 RANGE=100 %
10
O
2
range.
STABILITY
STABIL=0.0 PPM
3
CO STB=0.0 PPM
3, 7, 10
CO2 STB=0.0 %
7
O2 STB=0.0 %
10
Concentration stability).
RESPONSE
2
RSP=0.20(0.00) SEC
Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
COMEAS
CO MEAS=4125.0 MV
Detector measure reading.
COREF
CO REF=3750.0 MV
Detector reference reading.
MRRATIO
MR RATIO=1.100
Measure/reference ratio.
AUTOZERO
4, 5
AZERO RATIO=1.234
Measure/reference ratio during auto-
reference.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
PURGEPRESS
9
PURGE=7.5 PSIG
Purge pressure
VACUUM
8
VAC=6.8 IN-HG-A
Vacuum pressure.
SAMPFLOW
6
SAMP FL=751 CC/M
Sample flow rate.
SAMPTEMP
SAMPLE TEMP=26.8 C
Sample temperature.
BENCHTEMP
BENCH TEMP=48.1 C
Bench temperature.
WHEELTEMP
WHEEL TEMP=68.1 C
Wheel temperature.
O2CELLTEMP
10
O2 CELL TEMP=50.2 C
O
2
sensor cell temperature.
BOXTEMP
BOX TEMP=26.8 C
Internal box temperature.
BOXTEMP2
4
BOX TEMP2=29.6 C
Internal box temperature #2.
OVENTEMP
11
OVEN TEMP=30.1 C
Oven temperature
PHOTOTEMP
PHT DRIVE=2500.0 MV
Photometer temperature.
COSLOPE
SLOPE=1.000
CO SLOPE=1.000
7
CO slope for current range, computed
during zero/span calibration.
COSLOPE1
SLOPE1=1.000
CO SLOPE1=1.000
7
CO slope for range #1 in dual range
mode, computed during zero/span
calibration.
COSLOPE2
SLOPE2=1.000
CO SLOPE2=1.000
7
CO slope for range #2 in dual range
mode, computed during zero/span
calibration.
COOFFSET
OFFSET=0.000
CO OFFSET=0.000
7
CO offset for current range, computed
during zero/span calibration.
COOFFSET1
OFFSET1=0.000
CO OFFSET1=0.000
7
CO offset for range #1 in dual range
mode, computed during zero/span
calibration.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...