Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
104
5.8.
SETUP
à
VARS: VARIABLES SETUP AND DEFINITION
The T300/T300M has several user-adjustable software variables, which define certain
operational parameters. Usually, these variables are automatically set by the
instrument’s firmware, but can be manually redefined using the
VARS
menu.
The following table lists all variables that are available within the 818 password
protected level. See Appendix A-2 for a detailed listing of all of the T300/T300M
variables that are accessible through the remote interface.
Table 5-3: Variable Names (VARS)
NO.
VARIABLE
DESCRIPTION
ALLOWED
VALUES
VARS
DEFAULT
VALUES
0
DAS_HOLD_OFF
Changes the Internal Data Acquisition System (DAS)
HOLDOFF
timer (Section 7.1.11).
No data is stored in the DAS channels during situations
when the software considers the data to be questionable
such as during warm up of just after the instrument returns
from one of its calibration mode to
SAMPLE
Mode.
May be set for
intervals
between
0.5 – 20 min
15 min.
1
CONC
_
PRECISION
Allows the user to set the number of significant digits to the
right of the decimal point display of concentration and stability
values.
AUTO, 1, 2,
3, 4
AUTO
2
DYN_ZERO
1
Dynamic zero automatically adjusts offset and slope of the
CO response when performing a zero point calibration during
an AutoCal (see Section 9.4).
ON/OFF
OFF
3
DYN_SPAN
1
Dynamic span automatically adjusts the offsets and slopes of
the CO response when performing a slope calibration during
an AutoCal (see Section 9.4).
ON/OFF
OFF
4
CLOCK_ADJ
Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock is
too fast.
-60 to +60
s/day
0 sec
5
STABIL_GAS
2
Selects which gas measurement is displayed when the STABIL
test function is selected.
CO; CO
2
& O
2
CO
1
Use
of the DYN_ZERO and DYN_SPAN features are not allowed for applications requiring EPA equivalency.
2
This VARS only appears if either the optional O
2
or CO
2
sensors are installed.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...