Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
203
Only appears if
either the O2 or the
CO2 sensor is
installed.
Only appears if either the
O
2
or the CO
2
sensor is
installed.
SAMPLE
GAS TO CAL:CO
CO
O2
ENTR
EXIT
Press
ENTR
to change the
OFFSET
&
SLOPE
values for the CO
measurement.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ZERO CAL M
STABIL= XXXX PPM
CO=XXXX
<TST TST>
ZERO
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to change
the
OFFSET
&
SLOPE
values for the CO
measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the NO/CO
measurement.
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls
below
0.2 PPM (for T300)
or 1.0 PPM (for T300M).
This may take several
minutes.
ZERO CAL M
STABIL= XXXX PPM
CO=XXXX
<TST TST>
ENTR
CONC
EXIT
SAMPLE
GAS TO CAL:CO
CO
O2
ENTR
EXIT
SPAN CAL M
STABIL= XXXX PPM
CO=X.XXX
<TST TST> ZERO
SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls
below
0.2 PPM (for
T300) or 1.0 PPM (for
T300M).
This may take several
minutes.
SPAN CAL M
STABIL= XXXX PPM
CO=X.XXX
<TST TST>
ENTR
CONC
EXIT
SPAN CAL M
STABIL= XXXX PPM
CO=X.XXX
<TST TST> ENTR
CONC
EXIT
When the analyzer enters
SPAN
cal mode, the
SPAN
button appears.
You may see both buttons
during the transition from
ZERO to SPAN modes.
If either the
ZERO
or
SPAN
button fails to appear see
Section 11 for
troubleshooting tips.
Front panel display
simultaneously shows
CO, NO, and NO
2
measurements
throughout this
procedure.
SAMPLE RANGE=50.0PPM CO=XX.XX
< TST
TST >
CAL CALZ
CALS
SETUP
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST > CAL
CALZ
CALS
SETUP
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST > CAL CALZ
CALS
SETUP
Analyzers enters
ZERO
cal
mode
.
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST > CAL
CALZ CALS
SETUP
Only appears if the
DUAL
or
AUTO
range modes are
selected.
Use these buttons to
choose the appropriate
range.
Repeat entire procedure
for each range.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...