Troubleshooting and Service
Teledyne API – Model T300/T300M CO Analyzer
254
4. If gas flows through the instrument when it is disconnected from its sources of zero
air, span gas or sample gas, the flow problem is most likely not internal to the
analyzer. Check to make sure that:
•
All calibrators/generators are turned on and working correctly.
•
Gas bottles are not empty or low.
•
Valves, regulators and gas lines are not clogged or dirty.
11.2.2.2.
LOW FLOW
1. Check if the pump diaphragm is in good condition. If not, rebuild the pump (see
Section 10.3.2). Check the Spare Parts List for information on pump rebuild kits.
2. Check for leaks as described in Section 10.3.3. Repair the leaking fitting, line or
valve and re-check.
3. Check for the sample filter and the orifice filter for dirt. Replace filters (see 10.3.1).
4. Check for partially plugged pneumatic lines, or valves. Clean or replace them.
5. Check for plugged or dirty critical flow orifices. Replace them.
6. If an IZS option is installed in the instrument, press CALZ and CALS. If the flow
increases then suspect a bad sample/cal valve.
11.2.2.3.
HIGH FLOW
The most common cause of high flow is a leak in the sample flow control assembly or
between there and the pump. If no leaks or loose connections are found in the fittings or
the gas line between the orifice and the pump, replace the critical flow orifice(s) inside
the sample flow control assembly.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...