Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
221
9.7.2.2.
SET CO2 SPAN GAS CONCENTRATION:
Set the expected CO
2
span gas concentration.
This should be equal to the percent concentration of the CO
2
span gas of the selected
reporting range (default factory setting = 12%).
SAMPLE
GAS TO CAL:CO
CO
CO2
O2
ENTR EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
< TST TST >
CAL
SETUP
M-P CAL RANGE=50.0 PPM CO= XX.XX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
GAS TO CAL:O2
CO
CO2
O2
ENTR
EXIT
M-P CAL C O2 SPAN CONC:12.00%
0
1
2
.0
0
ENTR EXIT
The CO
2
span concentration value automatically defaults
to 12 %.
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the CO
2
calibration gases.
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the previous menu.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...