Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
218
9.7.1.3.
ACTIVATE O2 SENSOR STABILITY FUNCTION
To change the stability test function from CO concentration to the O
2
sensor output,
press:
Press
EXIT
3
times to return
to
SAMPLE
menu.
SETUP X.X
STABIL_GAS:O2
CO
CO2 O2
ENTR
EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
ALRM
EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
NEXT>
JUMP
EDIT PRNT EXIT
SETUP X.X
2) STABIL_GAS=CO
<PREV NEXT> JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS:CO
CO
CO2
O2
ENTR EXIT
Continue pressing
NEXT
until ...
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
CO2 and O2
options only
appear if
associated
sensors are
installed.
Note
Use the same procedure to reset the STB test function to CO when the O2
calibration procedure is complete.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...