Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
95
5.4.3.3.
AUTO RANGE MODE (AUTO)
In
AUTO
range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (low and high).
•
The unit will switch from low range to high range when the CO
2
concentration
exceeds 98% of the low range span.
•
The unit will return from high range back to low range once both the CO
2
concentration falls below 75% of the low range span.
In
AUTO
Range Mode the instrument reports the same data in the same range on both
the
A1
and
A2
outputs and automatically switches both outputs between ranges as
described above. Also the
RANGE
test function displayed on the front panel will be
replaced by two separate functions:
•
RANGE1
: The
LOW
range setting for all analog outputs.
•
RANGE2
: The
HIGH
range setting for all analog outputs.
The high/low range status is also reported through the external, digital status bits (See
Section 3.3.1.4).
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...