![TELEDYNE API T300 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/teledyne-api/t300/t300_operation-manual_1080880019.webp)
17
1. INTRODUCTION, FEATURES AND OPTIONS
This section provides an overview of Teledyne API’s Model T300 or T300M Analyzer,
its features and its options.
1.1.
T300 FAMILY OVERVIEW
The family includes the T300 and the T300M Gas Filter Correlation Carbon Monoxide
Analyzer. The T300 family of analyzers is a microprocessor-controlled analyzer that
determines the concentration of carbon monoxide (CO) in a sample gas drawn through
the instrument. It uses a method based on the Beer-Lambert law, an empirical
relationship that relates the absorption of light to the properties of the material through
which the light is traveling over a defined distance. In this case the light is infrared
radiation (IR) traveling through a sample chamber filled with gas bearing a varying
concentration of CO.
The T300/T300M uses Gas Filter Correlation (GFC) to overcome the interfering effects
of various other gases (such as water vapor) that also absorb IR. The analyzer passes the
IR beam through a spinning wheel made up of two separate chambers: one containing a
high concentration of CO known as the
reference
, and the other containing a neutral gas
known as the
measure
. The concentration of CO in the sample chamber is computed by
taking the ratio of the instantaneous measure and reference values and then
compensating the ratio for sample temperature and pressure.
The T300/T300M Analyzer’s multi-tasking software gives the ability to track and report
a large number of operational parameters in real time. These readings are compared to
diagnostic limits kept in the analyzers memory and should any fall outside of those
limits the analyzer issues automatic warnings.
Built-in data acquisition capability, using the analyzer's internal memory, allows the
logging of multiple parameters including averaged or instantaneous concentration
values, calibration data, and operating parameters such as pressure and flow rate. Stored
data are easily retrieved through the serial port or Ethernet port via our APICOM
software or from the front panel, allowing operators to perform predictive diagnostics
and enhanced data analysis by tracking parameter trends. Multiple averaging periods of
one minute to 365 days are available for over a period of one year.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...