Teledyne API – Model T300/T300M CO Analyzer
Theory of Operation
301
12.4.5.3.
THERMISTOR INTERFACE
This circuit provides excitation, termination and signal selection for several negative-
coefficient, thermistor temperature sensors located inside the analyzer. They are as
follows:
S
AMPLE
T
EMPERATURE
S
ENSOR
The source of this signal is a thermistor located inside the sample chamber of the Optical
Bench. It measures the temperature of the sample gas in the chamber. This data is used
to during the calculation of the CO concentration value.
B
ENCH
T
EMPERATURE
S
ENSOR
This thermistor is attached to the sample chamber housing. It reports the current
temperature of the chamber housing to the CPU as part of the bench heater control loop.
W
HEEL
T
EMPERATURE
S
ENSOR
This thermistor is attached to the heatsink on the GFC Wheel motor assembly. It reports
the current temperature of the wheel/motor assembly to the CPU as part of the Wheel
Heater control loop.
B
OX
T
EMPERATURE
S
ENSOR
A thermistor is attached to the motherboard. It measures the analyzer’s internal
temperature. This information is stored by the CPU and can be viewed by the user for
troubleshooting purposes via the front panel display (see Section 11.1.2).
12.4.5.4.
ANALOG OUTPUTS
The analyzer comes equipped with four analog outputs:
A1
,
A2
,
A3
and
A4
. The type
of data and electronic performance of these outputs are configurable by the user (see
Section 5.4).
O
UTPUT
L
OOP
-
BACK
All four analog outputs are connected back to the A/D converter through a loop-back
circuit. This permits the voltage outputs to be calibrated by the CPU without need for
any additional tools or fixtures.
12.4.5.5.
INTERNAL DIGITAL I/O
This channel is used to communicate digital status and control signals about the
operation of key components of the Optical Bench. The CPU sends signals to the
sync/demod board that initiate the
ELECTRICAL TEST
and
DARK CALIBRATION
procedures.
12.4.5.6.
EXTERNAL DIGITAL I/O
This External Digital I/O performs two functions: status outputs and control inputs.
S
TATUS
O
UTPUTS
Logic-Level voltages are output through an optically isolated 8-pin connector located on
the rear panel of the analyzer. These outputs convey good/bad and on/off information
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...