![TELEDYNE API T300 Скачать руководство пользователя страница 74](http://html.mh-extra.com/html/teledyne-api/t300/t300_operation-manual_1080880074.webp)
Getting Started
Teledyne API – Model T300/T300M CO Analyzer
72
3.4.4.1.
INTERFERENTS FOR CO MEASUREMENTS
It should be noted that the gas filter correlation method for detecting CO is subject to
interference from a number of other gases that absorb IR in a similar fashion to CO.
Most notable of these are water vapor, CO
2
, N
2
O (nitrous oxide) and CH
4
(methane).
The T300/T300M has been successfully tested for its ability to reject interference from
of these sources, however high concentrations of these gases can interfere with the
instrument’s ability to make low-level CO measurements.
For a more detailed discussion of this topic, see Section 12.2.1.3.
3.4.4.2.
INITIAL CALIBRATION PROCEDURE
The following procedure assumes that:
•
The instrument DOES NOT have any of the available calibration valve or gas inlet
options installed;
•
Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer
(see Figure 3-4)
•
The pneumatic setup matches that described in Section 3.3.2.1.
V
ERIFYING THE
T300/T300M
R
EPORTING
R
ANGE
S
ETTINGS
While it is possible to perform the following procedure with any range setting we
recommend that you perform this initial checkout using following reporting range
settings:
•
Unit of Measure: PPM
•
Analog Output Reporting Range: 50 ppm
•
Mode Setting: SNGL
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...