Theory of Operation
Teledyne API – Model T300/T300M CO Analyzer
280
12.1.
MEASUREMENT METHOD
This section presents measurement principles and fundamentals for this instrument.
12.1.1.
BEER’S LAW
The basic principle by which the analyzer works is called the Beer-Lambert Law or
Beer’s Law. It defines how light of a specific wavelength is absorbed by a particular gas
molecule over a certain distance. The mathematical relationship between these three
parameters is:
I = I
o
e
-
α
L
c
Equation 12-1
Where:
I
o
is the intensity of the light if there was no absorption.
I
is the intensity with absorption.
L
is the absorption path, or the distance the light travels as it is being absorbed.
C
is the concentration of the absorbing gas; in the case of the T300/T300M, Carbon
Monoxide (CO).
α
is the absorption coefficient that tells how well CO absorbs light at the specific
wavelength of interest.
12.2.
MEASUREMENT FUNDAMENTALS
In the most basic terms, the T300/T300M uses a high-energy heated element to generate
a beam of broad-band IR light with a known intensity (measured during instrument
calibration). This beam is directed through multi-pass cell filled with sample gas. The
sample cell uses mirrors at each end to reflect the IR beam back and forth through the
sample gas a number of times (see Figure 12-1).
The total length that the reflected light travels is directly related to the intended
sensitivity of the instrument. The lower the concentrations the instrument is designed to
detect, the longer the light path must be in order to create detectable levels of
attenuation.
Lengthening the absorption path is accomplished partly by making the physical
dimension of the reaction cell longer, but primarily by adding extra passes back and
forth along the length of the chamber.
Table 12-1: Absorption Path Lengths for T300 and T300M
MODEL
TOTAL NUMBER OF
REFLECTIVE PASSES
DISTANCE BETWEEN MIRRORS
TOTAL
ABSORPTION LIGHT PATH
T300
32
437.5 mm
14 Meters
T300M
8
312.5 mm
2.5 Meters
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...