Timing Parameters
1-58
DTG5078 & DTG5274 & DTG5334 Data Timing Generator User Manual 2
Operation.
You can set up Cross Point only in a PW/Duty/Cross Point cell for either of the
following channels:
A channel has not been assigned and its format is NRZ
A channel of the DTGM30, DTGM31 or DTGM32 installed in Slots A to D
has been assigned and its format is NRZ
If you use Cross Point for the setting, the value will be prefixed with the letter “C”.
Long Delay
For each Delay and Pulse Width, the variable range supported by the
DTG5000
Series
hardware is only approximately 5 ns. For greater amounts of Delay and Pulse
Width, 480 ns is enabled at the maximum by internally manipulating the data
pattern by software.
Long Delay writes the pattern into the pattern memory while displacing it by
software.
Figure 1-34: Writing the pattern into the memory while displacing it
Displacement of the pattern allows you to set the value only with an integral
multiple of the clock frequency. Therefore, the section under the clock frequency
is required to be delayed by hardware.
If the clock frequency is later than the valid range of hardware delay, you cannot
set the Delay value in a certain range. To avoid this problem, the software enlarges
the pattern.
For example, if you want to output a “010” pattern with a clock frequency of
200 MHz, you cannot set the Delay to 9 ns. This is because the hardware delay can
be up to 5 ns. In this case, the software writes a double-length enlarged pattern
internally into the pattern memory to operate the clock at the double rate.
0
1
0
0
1
1
0
0
10ns
0
0
1
0
0
1
1
0
User-defined pattern
Contents of pattern memory
When a clock frequency is 100MHz
Содержание DTG5078
Страница 4: ......
Страница 10: ...Table of Contents iv DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 19: ...Reference...
Страница 20: ......
Страница 24: ...Operating Modes 1 4 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 34: ...Data Structure 1 14 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 58: ...Pattern Edit 1 38 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 88: ...Timing Parameters 1 68 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 100: ...DC Output 1 80 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 120: ...Jitter Generation 1 100 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 126: ...DTG5000 Configuration Utility 1 106 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 146: ...Pulse Generator mode 1 126 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 150: ...Offline mode 1 130 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 176: ...Diagnostics and Calibration 1 156 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 177: ...Appendices...
Страница 178: ......
Страница 186: ...Appendix B System Recovery B 4 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 189: ...Index...
Страница 190: ......
Страница 196: ...Index Index 6 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...