Section 2—2445 Service
OPERATING INFORMATION
This section of the manual provides information on instrument installation and power requirements; and the functions of
controls, connectors, and indicators are described. Operating considerations, intended to familiarize the operator w ith basic
measurement techniques, and operator's checks and adjustments fo r the 2445 are included. For additional operating
information, refer to the 2445 Operators Manual.
PREPARATION FOR USE
SAFETY CONSIDERATIONS
Refer to the Safety Summaries at the fro n t of this
manual fo r power source, grounding, and other safety
considerations pertaining to the use of the instrument.
Before connecting the oscilloscope to a power source,
read entirely both this section and the Safety Summaries.
This instrument may be damaged i f operated with
the LIN E VOLTAGE SELECTOR switch set fo r the
wrong applied ac input-source voltage o r i f the
wrong line fuse is installed.
LINE VOLTAGE SELECTION
The 2445 operates from either a 115-V or a 230-V
nominal ac power-input source having line frequency
ranging from 48 Hz to 440 Hz. Before connecting the
power cord to a power-input source, verify that the LINE
VOLTAGE SELECTOR switch, located on the rear panel
(see Figure 2-1), is set fo r the correct nominal ac input-
source voltage. To convert the instrument fo r operation
from one line-voltage range to the other, move the LINE
VOLTAGE SELECTOR switch to the correct nominal
ac source-voltage setting (see Table 2-1). The detachable
power cord may have to be changed to match the particular
power-source output.
LINE FUSE
To verify that the instrument power-input fuse is of
proper value for the nominal ac source voltage selected,
perform the following procedure:
1. Press in the fuse-holder cap and release it w ith a
slight counterclockwise rotation.
2. Pull the cap (with the attached fuse inside) out of
the fuse holder.
3. Verify proper fuse value (see Table 2-1).
4. Install the proper fuse and reinstall the fuse-holder
cap.
NOTE
The two types o f fuses listed are n o t directly inter
changeable; they require different types o f fuse caps
POWER CORD
This instrument has a detachable, three-wire power cord
w ith a three-contact plug fo r connection to both the power
source and protective ground. The power cord is secured to
the rear panel by a cord-set-securing clamp. The protective-
ground contact on the plug connects (through the power-
cord protective grounding conductor) to the accessible
metal parts of the instrument. For electrical-shock pro
tection, insert this plug into a power-source outlet that has
a properly grounded protective-ground contact.
Instruments are shipped w ith the required power cord
as ordered by the customer. Available power-cord infor
mation is presented in Table 2-1. Part numbers are listed
both in the "Accessories" information at the rear of this
manual and in the "O ptions and Accessories" section of
the Operators Manual (Section 7). Contact your Tektronix
representative or local Tektronix Field Office fo r additional
power-cord information.
2-1
Содержание 2445
Страница 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Страница 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Страница 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Страница 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Страница 210: ...3829 58 Figure 9 4 2445 block diagram ...
Страница 214: ......
Страница 217: ......
Страница 219: ...2445 382 72 ...
Страница 222: ...2445 ...
Страница 231: ...A 1 t C t t F t G t H t ...
Страница 233: ......
Страница 236: ......
Страница 238: ...2445 392 1 75 ...
Страница 244: ......
Страница 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Страница 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Страница 253: ......
Страница 263: ... 0 2445 J8 i S ...
Страница 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Страница 275: ......
Страница 278: ......
Страница 281: ......
Страница 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Страница 286: ...2445 3823 82 ...
Страница 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Страница 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Страница 300: ......
Страница 301: ...J119 O A A Z fio rv ic A J118 VERT READOUT JITTER R618 i x n n t o i n R801 X10 GAIN R850 3829 71 ...
Страница 304: ...2 R E TU R N T O 1 ...
Страница 305: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 306: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Страница 308: ...RETURN TO ...
Страница 309: ...FRONT PANEL TROUBLESHOOTING ...
Страница 310: ...FRONT PANEL TROUBLESHOOTING ...
Страница 311: ...2445 Service 3829 90 ...
Страница 316: ...R E TU R N T O v 1 y ...
Страница 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Страница 318: ...2445 Service NOTE 3 56V IS M E A S U R E D A S 1 44V W ITH R ESPEC T TO TH E 5V SUPPLY 3829 95 ...
Страница 323: ......
Страница 324: ...2445 Service 3829 85 ...
Страница 325: ......
Страница 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Страница 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Страница 330: ......
Страница 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Страница 338: ......
Страница 346: ...12 2445 SERVICE ...
Страница 347: ...2445 SERVICE ...
Страница 348: ...2445 SERVICE ...