Maintenance— 2445 Service
4. Clean switches with isopropyl alcohol and wait 60 sec
onds for the majority of the alcohol to evaporate. Then com
plete drying with low-pressure air.
To p reven t dam age from electrical arcing, ensure th at
circuit boards an d com ponents a re d ry before apply
ing p o w e r to the instrum ent.
5. Dry all components and assemblies in an oven or dry
ing compartment using low-temperature (125° F to 150° F)
circulating air.
CLEANING. To clean the interior, blow off dust with dry,
low-pressure air (approximately 9 psi). Remove any remain
ing dust with a soft brush or a cloth dampened with a solu
tion of mild detergent and water. A cotton-tipped applicator
is useful for cleaning in narrow spaces and on circuit boards.
LUBRICATION
If these methods do not remove all the dust or dirt, the
instrument may be spray washed using a solution of 5%
mild detergent and 95% water as follows:
There is no periodic lubrication required for this
instrument.
SEMICONDUCTOR CHECKS
E xceptions to the follow ing procedure are the A tten u
a to r assem blies an d the F ro nt-P anel m odule. C lean
these assem blies only w ith isopropyl alcohol as de
scribed in step 4.
Periodic checks of the transistors and other semiconduc
tors in the oscilloscope are not recommended. The best
check of semiconductor performance is actual operation in
the instrument.
1. Gain access to the parts to be cleaned by removing
easily accessible shields and panels.
PERIODIC READJUSTMENT
2. Spray wash dirty parts with the detergent-and-water
solution; then use clean water to thoroughly rinse them.
To ensure accurate measurements, check the perfor
mance of this instrument every 2000 hours of operation, or
if used infrequently, once each year. In addition, replace
ment of components may necessitate readjustment of the
3. Dry all parts with low-pressure air.
affected circuits.
N O T E
Complete Performance Check and Adjustment instruc
M o s t o f the sw itches used in the 2 4 4 5 a re sealed an d
the contacts a re inaccessible. If cleaning is deem ed
necessary, use only isopropyl alcohol.
tions are given in Sections 4 and 5. The Performance Check
Procedure can also be helpful in localizing certain troubles in
the instrument.
6-4
Содержание 2445
Страница 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Страница 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Страница 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Страница 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Страница 210: ...3829 58 Figure 9 4 2445 block diagram ...
Страница 214: ......
Страница 217: ......
Страница 219: ...2445 382 72 ...
Страница 222: ...2445 ...
Страница 231: ...A 1 t C t t F t G t H t ...
Страница 233: ......
Страница 236: ......
Страница 238: ...2445 392 1 75 ...
Страница 244: ......
Страница 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Страница 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Страница 253: ......
Страница 263: ... 0 2445 J8 i S ...
Страница 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Страница 275: ......
Страница 278: ......
Страница 281: ......
Страница 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Страница 286: ...2445 3823 82 ...
Страница 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Страница 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Страница 300: ......
Страница 301: ...J119 O A A Z fio rv ic A J118 VERT READOUT JITTER R618 i x n n t o i n R801 X10 GAIN R850 3829 71 ...
Страница 304: ...2 R E TU R N T O 1 ...
Страница 305: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 306: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Страница 308: ...RETURN TO ...
Страница 309: ...FRONT PANEL TROUBLESHOOTING ...
Страница 310: ...FRONT PANEL TROUBLESHOOTING ...
Страница 311: ...2445 Service 3829 90 ...
Страница 316: ...R E TU R N T O v 1 y ...
Страница 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Страница 318: ...2445 Service NOTE 3 56V IS M E A S U R E D A S 1 44V W ITH R ESPEC T TO TH E 5V SUPPLY 3829 95 ...
Страница 323: ......
Страница 324: ...2445 Service 3829 85 ...
Страница 325: ......
Страница 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Страница 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Страница 330: ......
Страница 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Страница 338: ......
Страница 346: ...12 2445 SERVICE ...
Страница 347: ...2445 SERVICE ...
Страница 348: ...2445 SERVICE ...