Operating Information—2445 Service
CONTROLS, CONNECTORS,
AND INDICATORS
The follow ing descriptions are intended to familiarize
the operator w ith the location and function o f the instru
ment's controls, connectors, and indicators.
POWER AND DISPLAY
Refer to Figure 2-2 fo r location of items 1 through 9.
0
INTENSITY Control—Adjusts brightness o f the crt
trace display. This control does not affect intensity of
the c rt readout display.
BEAM FIND Switch—When held in, compresses the
display to w ithin the graticule area. Aids the operator
in locating off-screen displays.
0
FOCUS Control—Adjusts the display for optimum
definition.
^ 4 ^ TRACE
ROTATION
Control—Operator-adjusted
screwdriver control used to align the crt trace w ith
the horizontal graticule lines. Once adjusted, it does
not require readjustment during normal operation of
the instrument.
©
READOUT INTENSITY Control—Adjusts the inten
sity of the crt readout display. This control is also
used to either enable or disable the scale-factor
display. Setting the control to MIN reduces the
readout intensity to minimum. Clockwise rotation
from midrange increases the readout intensity and
enables the scale-factor display; counterclockwise
rotation from midrange also increases the intensity
but disables the scale-factor display. Delta Volts and
Delta Time readouts and control messages w ill
continue to be enabled even when the scale-factor
display is disabled.
ASTIG Control—Operator-adjusted screwdriver con
trol used in conjunction w ith the FOCUS control to
obtain a well-defined display over the entire graticule
area. Once adjusted, it does not require readjustment
during normal operation of the instrument.
( T ) SCALE ILLU M C ontrol-Adjusts the light level o f
the graticule illumination.
( jT ) POWER Switch—Turns instrument power on and off.
Press in for ON; press again for OFF. An internal
indicator in the switch shows green when the switch
is on and black when it is o ff. Front-panel settings
that were unchanged fo r at least 10 seconds prior to
power-off w ill be returned when power is reapplied to
the instrument.
0
CRT —Has an 80-mm vertical and 100-mm horizontal
display area. Internal graticule lines eliminate parallax
viewing error between the trace and the graticule
lines. Rise-time measurement points are indicated
at the left edge o f the graticule.
2-4
Содержание 2445
Страница 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Страница 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Страница 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Страница 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Страница 210: ...3829 58 Figure 9 4 2445 block diagram ...
Страница 214: ......
Страница 217: ......
Страница 219: ...2445 382 72 ...
Страница 222: ...2445 ...
Страница 231: ...A 1 t C t t F t G t H t ...
Страница 233: ......
Страница 236: ......
Страница 238: ...2445 392 1 75 ...
Страница 244: ......
Страница 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Страница 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Страница 253: ......
Страница 263: ... 0 2445 J8 i S ...
Страница 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Страница 275: ......
Страница 278: ......
Страница 281: ......
Страница 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Страница 286: ...2445 3823 82 ...
Страница 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Страница 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Страница 300: ......
Страница 301: ...J119 O A A Z fio rv ic A J118 VERT READOUT JITTER R618 i x n n t o i n R801 X10 GAIN R850 3829 71 ...
Страница 304: ...2 R E TU R N T O 1 ...
Страница 305: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 306: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Страница 308: ...RETURN TO ...
Страница 309: ...FRONT PANEL TROUBLESHOOTING ...
Страница 310: ...FRONT PANEL TROUBLESHOOTING ...
Страница 311: ...2445 Service 3829 90 ...
Страница 316: ...R E TU R N T O v 1 y ...
Страница 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Страница 318: ...2445 Service NOTE 3 56V IS M E A S U R E D A S 1 44V W ITH R ESPEC T TO TH E 5V SUPPLY 3829 95 ...
Страница 323: ......
Страница 324: ...2445 Service 3829 85 ...
Страница 325: ......
Страница 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Страница 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Страница 330: ......
Страница 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Страница 338: ......
Страница 346: ...12 2445 SERVICE ...
Страница 347: ...2445 SERVICE ...
Страница 348: ...2445 SERVICE ...