Adjustment Procedure— 2445 Service
2. Check/Adjust CRT Termination (R1501),
Vertical Gain (R638), Vertical Centering (R639),
High-Frequency Trans Resp (C404, R403), Vertical
Readout Jitter (R618), Horizontal Readout Jitter
(R805), and X-Y Phasing (C l 18).
N OTE
If the previous step was not performed, first setup the
Initial Control Settings before the Dynamic Centering
adjustments, then proceed as follows.
C R T
Termination,
High
Frequency
Transient
Response, Vertical Gain, Vertical Centering, and
Readout Jitter adjustments are interactive. This pro
cedure optimizes these adjustments together.
a. Rotate the AREF OR DLY POS control CCW until the
RATIO readout is constant.
N O TE
As a guide when performing the following adjust
ments, optimum performance is achieved when the
CH 1 and CH 2 step response aberrations are
<
4%
when using 10 mV/division deflection factors
0.2
division on a 5-division signal).
i.
ADJUST—CRT Termination (R1501) for best flat-top
approximately 5 ns past the rising edge of the waveform.
The adjustment is accessable through a hole in the top
cover plate. Squeezing the output leads of the termination
inductors (LR1513 and LR1514) toward each other will re
duce the spike that may be present approximately 6 ns be
hind the leading edge.
j.
ADJUST—Trans Resp adjustments (C404 and R403)
alternately for the best flat top on the first 10 ns of the
positive portion of the waveform.
b. Rotate the A control until the readout display indi
cates 130.0%.
c.
CHECK— One cursor is near the bottom horizontal
graticule line and the other is near dotted graticule line
marked 100(%).
d.
Rotate the AREF OR DLY POS control until the read
out displays exactly 100.0%. The cursors should now be on
or near the dotted graticule lines marked 0% and 100(%).
e. Set the CH 1 VOLTS/DIV VAR to the detent position.
N O TE
Care must be taken not to disturb the position o f the
controls adjusted in parts b through e during the bal
ance o f this procedure. If they are accidentally moved,
repeat the procedure from the beginning.
k.
ADJUST—Vertical Gain (R638) and Vertical Center
ing (R639) to vertically center the cursors precisely 5 divi
sions apart (align with the dotted 0% and 100(%) graticule
lines).
I. Press the AV button to turn off the cursors.
m.
Move the test signal to CH 2 and set the VERTICAL
MODE switches to display CH 2.
n.
Repeat part j for CH 2, switching between
CH 1 and CH 2 as necessary, until both CH 2 and CH 1
aberrations are minimized. When minimized, leave CH 2
selected.
o.
Disconnect the Calibration Generator and connect the
Leveled Sine-Wave Generator to the CH 2 input via a BNC
cable.
f. Connect the high-amplitude output of the Calibration
p. Set the generator for a 6-division display at the refer-
Generator to the CH 1 OR X input connector via a 50 S2 BNC
ence frequency,
cable, a Tunnel Diode Pulssr, and a 5X attenuator.
g.
Set the generator Period switch to 100 kHz, and set
the generator amplitude control to maximum.
h.
Rotate the pulser Trigger control CW (from a fully
CCW position) until a stable signal first appears on the grati
cule. Display amplitude will be approximately 5 divisions.
The oscilloscope TRIGGER LEVEL control may need to be
adjusted to obtain a stable display.
q. Change the generator output frequency to 150 MHz.
/
-
'
L
l
a a
i . v /i i L v r \ — L / i o p i a y a m p i i i u u ^ io u c i w c c i i *t.*t u i v o i u i i o
and 5 divisions. This bandwidth provides optimum perfor
mance of the Vertical system.
s.
ADJUST— If necessary, compromise the settings in
part j to obtain the best flat top with the proper bandwidth.
5-17
Содержание 2445
Страница 1: ...Tektronix 2445 OSCILLOSCOPE SERVICE INSTRUCTION MANUAL ...
Страница 11: ...2445 Service 3829 01 The 2445 Oscilloscope ...
Страница 44: ...Theory of Operation 2445 Service 3831 10A Figure 3 1 Block diagram ...
Страница 45: ...Theory of Operation 2445 Service 3831 10B Figure 3 1 Block diagram cont 3 3 ...
Страница 210: ...3829 58 Figure 9 4 2445 block diagram ...
Страница 214: ......
Страница 217: ......
Страница 219: ...2445 382 72 ...
Страница 222: ...2445 ...
Страница 231: ...A 1 t C t t F t G t H t ...
Страница 233: ......
Страница 236: ......
Страница 238: ...2445 392 1 75 ...
Страница 244: ......
Страница 247: ...A 1 C _____ D E F G H J 2445 3811 74 ...
Страница 248: ...1 2 3 4 5 6 7 8 9 i o 2445 DISPLAY SEQUENCER TRIG GERING A4B SWEEPS ...
Страница 253: ......
Страница 263: ... 0 2445 J8 i S ...
Страница 264: ...1 2 3 4 5 6 7 i 8 I i 9 10 2445 READOUT ...
Страница 275: ......
Страница 278: ......
Страница 281: ......
Страница 283: ... 8VJNR EG 3S 5 fROM P232 5 10 A 15VUNREG 8S F R O Mn i 2445 3 0 2 S 8 I ...
Страница 286: ...2445 3823 82 ...
Страница 290: ...B H le w o q 87V T S o I R v n i U1 R1873 PARTIAL A9 HIGH VOLTAGE BOARD 2445 ...
Страница 299: ...2445 Service DAC REF A5 CONTROL ADJUSTMENT LOCATIONS 3 ...
Страница 300: ......
Страница 301: ...J119 O A A Z fio rv ic A J118 VERT READOUT JITTER R618 i x n n t o i n R801 X10 GAIN R850 3829 71 ...
Страница 304: ...2 R E TU R N T O 1 ...
Страница 305: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 306: ...ERROR MESSAGE DIAGNOSTICS ...
Страница 307: ...O A A C t rnra g i tiw c t 3829 89 ...
Страница 308: ...RETURN TO ...
Страница 309: ...FRONT PANEL TROUBLESHOOTING ...
Страница 310: ...FRONT PANEL TROUBLESHOOTING ...
Страница 311: ...2445 Service 3829 90 ...
Страница 316: ...R E TU R N T O v 1 y ...
Страница 317: ...SWEEP TROUBLESHOOTING PROCEDURE ...
Страница 318: ...2445 Service NOTE 3 56V IS M E A S U R E D A S 1 44V W ITH R ESPEC T TO TH E 5V SUPPLY 3829 95 ...
Страница 323: ......
Страница 324: ...2445 Service 3829 85 ...
Страница 325: ......
Страница 326: ... KERNEL NOP DIAGNOSTIC PROCEDURE ...
Страница 327: ...10 POWER SUPPLY TROUBLESHOOTING PROCEDURE 3829 94 ...
Страница 330: ......
Страница 334: ...2445 Service REGULATOR TROUBLESHOOTING PROCEDURE 3829 93 ...
Страница 338: ......
Страница 346: ...12 2445 SERVICE ...
Страница 347: ...2445 SERVICE ...
Страница 348: ...2445 SERVICE ...