Basic Operations
------------------------------------------------------------------------------------
37
Error Codes
Position
Item
Number
Display
Top
Number
Display
Bottom
Cause
Verify/Recovery
WDT
E001
-
CPU may have performed an
illegal operation
Turn POWER switch OFF and restart
AD_OV
E002
-
Abnormal voltage detected in
the power supply circuit
May be damaged.
Contact Takasago Sales Dept.
OHP
E003
OHP
Over heating protection
is triggered.
●Turn power OFF. Leave device to cool
down. Turn power ON
●Check environmental temperatures
●Verify Fan Motor rotates when power is
turned on. (OUTPUT is OFF)
OVP
E004
OUP
Over voltage protection was
triggered
Check OVP values.
See page 30 to cancel alarm.
OCP
E005
OCP
Over current protection was
triggered
Check OCP values
See page 32 to cancel alarm
BST
E006
-
Slave unit malfunctioned.
(Displays only on Master
when in Parallel operation)
Check slave unit display
P_ERR
E007
-
May have May have had a
short interruption or voltage
dip at the main power
Turn power switch OFF. Check
device’s power supply.
BST_NRDY
E008
-
Slave unit's input voltage is
abnormal or POWER switch is
OFF
Check whether slave unit’s power
switch is ON
Device Detection
Failure
E009
-
Cannot detect number of
parallel devices
Restart power supply
Max. Device No.
OVER
E010
-
More than 11 parallel devices
are connected
Verify No. of parallel devices.
Restart power supply
Device No.
Disagreement
E011
-
Difference in the current No.
of connected parallel devices
and the previous No. of
parallel devices remembered
by the ZX power supply
If the present number of parallel
devices is satisfactory, hold the FUNC.
key for more than 2 seconds and
confirm the detected number of
devices.
TRIP
(1600LN/LAN)
※Displayed briefly
E012
Shdn
TRIP function was triggered
Cancel by pressing OVP + OCP keys
simultaneously or by the remote control
command ALARM CLEAR.
Fan Alarm
E013
FAn
Detected fan alarm.
Cancel by pressing OVP + OCP keys
simultaneously or by the remote control
command ALARM CLEAR.
The deterioration of the motor or the
trouble is possible.
(This function is only 400LN/LAN)
Содержание ZX-S-LN Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...