Calibration
------------------------------------------------------------------------------------
160
7. Press the PRESET key ①. Set to “80.00V” with
the VOLTAGE dial ②.
(See Operating as Constant Voltage Power Supply
under Basic Operations on how.)
8.Press the PRESET key.
9. Press the FUNC. key and change the top Number
Display to Settings Item 21 with the VOLTAGE dial.
10 . Adjust so that the digital multimeter (used for
output voltage measurement) displays “80.000V”
with the CURRENT dial.
11. Press the FUNC. key to complete full-scale
calibration of the output voltage settings.
12 . This completes offset/full-scale calibration of
Output Voltage Settings.
Please return the voltage/current set values back
to normal before use.
FUNC. lamp will light and the display on the bottom Number
Display will light out.
MEMO
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Be certain to perform this along with meter calibration
(Offset/Full-Scale Calibration of Voltmeter.)
Caution
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
ZX-S-4 00 LN
Содержание ZX-S-LN Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...