Using via Digital Data Communication
------------------------------------------------------------------------------------
115
SET Commands
This section explains the SET commands of EX based commands.
A
:Assigns Device Address
Function:Assigns the device address for ZX power supplies
Format:A* *:Set value within range
Range:0~50
Assigned Address 0 is a global address. Assign addresses to all multi-connected ZX-S Series power
supplies.
Once the global address is assigned, only Output ON/OFF Toggle SET command “OT” is valid.
Further, no Acknowledge Responses will be sent only while the global address is assigned
<Ex.>A1 /* Assigns address 1 to ZX power supply */
<Ex.>A1,OT1,A2,OT1 /* Errors due multple address assignment commands */
AR
:Executes Alarm Reset
Function:Executes alarm reset (cancels triggered OVP/OCP alarms)
Format:AR* *:Set value
Set value:
1:Resets alarms
Values other than 0 and 1 result in parameter errors. (0 does nothing)
<Ex.>A1,AR1
CL
:Resets Set Parameters
Function:Resest the set parameters
Format:CL* *:Set value
Set Value:
1:Resest to factory default settings
Values other than 0 and 1 result in parameter errors. (0 does nothing)
<Ex.>A1,CL1
Do not set redundant ZX power supply addresses within the same system.
Further, only one address can be assigned per send. Assigning multiple addresses at once is not
possible.
Caution
Execute alarm reset only once cause has been resolved.
Caution
Содержание ZX-S-LN Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...